Thermo Fisher SOLA II- Sulfur On-Line Analyzer
 
Specifications
 
Analytical Performance
  • Detector: Pulsed UV Fluorescence (PUVF) with Pyrolyzer for Total Sulfur Measurement1
  • Measuring Ranges : 
  • SOLA II : Full scale ranges from 0-5 ppm S to 0-5000 ppm S 
  • SOLA Trace: Full scale ranges from 0-2 ppm S to 0 - 5 ppm S                                                 
  • SOLA Flare DI (low and high range analyzer system): 10 ppm to 95% S by volume
  • SOLA Dual Calibration/Dual Stream: Same ranges available as on the SOLA II
  • Repeatability:  SOLA II: ±2% of full scale, one sample injection per minute2; ±1% of full scale, two sample injections per minute2
SOLA Trace: Consult Thermo Fisher
SOLA Flare DI (low and high range analyzer system): ±1% of full scale
SOLA Dual Calibration/Dual Stream: Refer to SOLA II
  • Linearity: ±2% of full scale, one sample injection per minute2; ±1% of full scale, two sample injections per minute2
  • Response Time: Semi-continuous, outputs updated every 1 second, 3-5 minutes to 90% of new value
  • Number of Process Streams: Dual streams with auto stream select (optional)
  • Calibration/Validation: Automatic or manual