- Analyzer : High-Energy Scanning Magnetic Sector
- Area Classification: General Purpose, or ATEX Zone 1, or Class 1 Div 2 (Z-purged) or Class 1 Div1 (X-purged)
- Dynamic Range: Faraday Detector: 10ppm - 100% (Application Dependent) Dual Faraday/SEM: 10ppb - 100% (Application Dependent)
- Analysis Time: 1-30 Seconds (Application Dependent)
- Analytical Precision: Typically 0.1% Relative Standard Deviation (24 Hour) for concentrations in the range 1-100%
- Calibration Interval: Typically 3-12 Months (Application Dependent)
- Service Interval: Analyzer 3 years, External Vacuum Pump 6-12 months(Application Dependent)
- Dimensions: 0.97m Width x 0.72m Depth x 1.48m Height, 240 kg approximate weight
- Power: 115 VAC (±5VAC) or 230 VAC (±10VAC) Consumption 2.0-2.5 kVA
- Environmental: Ideally 20°-25°C. Allowable Range between 12°C and 40°C
- Inlet Options Hazardous Area:
64-Port RMS (integral to cabinet or remote)
Dual RMS with 64-Port RMS integral to cabinet + remote 32 or 64 Port RMS
Single-Port Variable Pressure (0.1-1,000 mbar)
6-Port Variable Pressure (0.1-1,000 mbar)
General Purpose - all of the above plus:
Single Sample Port + 6 Calibration Ports
6 Sample Ports + 6 Calibration Ports
Batch Inlet Reservoir System
- Calibration Ports: Maximum 24 with single RMS or 48 with Dual RMS
- Standard Communications:4-Ports (Modbus (serial RS232, RS422 or RS485) and/or OPC), 13 Digital Outputs, 13 Digital Inputs
- Additionally Available Communication Protocols
Daniel Modbus, Echelon Lonworks, GE Fanuc Genius, MicroTrac Arcnet
Modicon (Modbus, Modbus Plus, Remote I/O), Profibus-DP,Reliance Automate